RT721 Programmable RTD Temperature Transmitter
Brand: Intempco Controls, IncThe RT721 is a low cost three-wire transmitter that converts a Pt100 ohm input to proportional voltage signal.
The transmitter provides sensor excitation and includes linearization, lead wire compensation, and lead-break detection functions.
Setup and calibration are made with a USB connection to your PC and Intempco’s configuration software.
Advanced signal processing capabilities and variable range input make this instrument very suitable for the most demanding temperature measurement applications.
Features- Easy setup, digital calibration and fully field re-programmable via USB module and PC Windows based software
- Factory calibrated or customer calibrated
- 0-1 Vdc, 0-5 Vdc, 1-5 Vdc or 0-10 Vdc output
- High accuracy and linearity Input 100 ohm Pt RTD (385 alpha) with 3-wire compensation
- Low temperature drift and wide ambient temperature range
- Metal housing mounts in DIN form B sensor heads
Specifications
Brand
Accuracy
- ± (0.05° C (0.09° F) + 0.05 % of Calibrated Span) or Better with Two-Point Calibration with Sensor
- ± 0.10 % of Span (Includes the Effects of Linearization & Repeatability, But Not Include Sensor Error)
Adjustment Type
- Factory Set (Pre-Set) or Customer Calibrated
Communication
- USB
Current
- Excitation Current RTD: 0.2 mA
Enclosure Finish
- Enamel Painted
Enclosure, Body Material
- Die Cast Zinc
Mounting
- DIN Rail (DIN 46277)
Operating Temperature
- -40° to 80° C (-40° to 176° F)
Output
- 0-1 Vdc
- 0-10 Vdc
- 0-5 Vdc
- 10 mA - Max.
- 1-5 Vdc
Power Supply
- 12-32 Vdc, Polarity Protected
Power Supply Effect
- < 0.001 %/V
Resistance
- Sensor Lead Resistance RTD: RTD Resistance + 2 times lead wire resistance must be less than 4 kiloohms
Resolution
- Output Resolution: 0.16 mV (15-bit)
RFI/EMI Effect
- 1 % or Less Typical
RTD Type
- Pt100, 3-Wire, Alpha = 0.00385, DIN EN 60751
Sensor, Probe Connection
- 3-Wire
Sensor, Probe Type
- RTD
Span - Minimum
- 50° C (122° F)
Span & Zero Adjustments
- By Software
Stability
- Long Term Drift: ≤ 0.1 % of Full Scale Per Year
Storage Temperature
- -40° to 80° C (-40° to 176° F)
Temperature Effect
- ± 0.002° C/°C
Temperature Range
- Software Re-Scalable between -200° to 600° C (-328° to 1,112° F)
Update Time
- Warmup: 30 seconds