Accuracy
- ± (0.10° C (0.18° F) + 0.10 % of Calibrated Span) with Two-Point Calibration
- ± (0.25° C (0.45° F) + 0.40 % of Span) with One-Point Calibration
Approvals & Certifications
Cable Material
- PVC Insulation
- Teflon Insulation 200° C (392° F), with Stainless Steel Armour & Polyolefin Shrink 90° C (195° F), with Stainless Steel Armour & Teflon Shrink 200° C (392° F) - Max.
Diameter
- Probe Diameter: 0.125 to 0.25” (3.175 to 6.35 mm)
Environmental Protection
- IP67 Hermetically Sealed
- NEMA 6P Hermetically Sealed
Input Impedance
- 1,000 ohms Min. (of Measuring Device, for Voltage Output)
Operating Pressure
- Sensor psig: 500 (kPa: 3,447, bar: 34.47)
Operating Temperature
- Housing: -40° to 80° C (-40° to 176° F)
Output
- 0-10 Vdc Linear to Temperature
- 0-5 Vdc Linear to Temperature
- 1-5 Vdc Linear to Temperature
- 4-20 mA Loop Powered, Linear to Temperature
Power Supply
- 12-32 Vdc Polarity Protected
Sensor Circuit
- 2.5 mA Downscale for 4-20 mA Output
- 24 mA Upscale for 4-20 mA Output
Sensor, Probe Material
- 316 Stainless Steel
- 316L Stainless Steel
Stability
- Long Term Drift: ≤ 0.1 % of Full Scale Per Year
Storage Temperature
- -40° to 80° C (-40° to 176° F)
Temperature Range
- Sensor: -200° to 600° C (-328° to 1,112° F)
Tri-Clamp Connection
- 0.5” Hygienic Clamp Union
- 0.75” Hygienic Clamp Union
Wetted Materials
- 316 Stainless Steel
- 316L Stainless Steel